[ CARE AND USE MANUAL ]

7

XSelect CSH HPLC Columns

There are two ways to fix the problem:

1. Tighten the screw a bit more. The ferrule moves forward, and reaches

the sealing surface. Do not over tighten since this may end in breaking
the screw.

2. Cut the tubing, replace the ferrule and make a new connection.

Alternatively, replace the conventional compression screw fitting
with an all-in-one PEEK fitting (Waters part number PSL613315)
that allows resetting of the ferrule depth. Another approach is to
use a SLIPFREE

®

connector to always ensure the correct fit. The

fingertight SLIPFREE

®

connectors automatically adjust to fit all

compression screw type fittings without the use of tools (Figure 5).

Figure 5: Single and Double SLIPFREE

®

Connectors

SLIPFREE

®

Connector Features:

• Tubing pushed into endfitting, thereby guaranteeing a void-free

connection

• Connector(s) come(s) installed on tubing

• Various tubing IDs and lengths available

• Fingertight to 10,000 psi – never needs wrenches

• Readjusts to all column endfittings

• Compatible with all commercially available endfittings

• Unique design separates tube-holding function from sealing function

Table 5: Waters Part Numbers for SLIPFREE

®

Connectors

Band Spreading Minimization

Figure 6 shows the influence of tubing internal diameter on system
band spreading and peak shape. As can be seen, the larger tubing
diameter causes excessive peak broadening and lower sensitivity.

Figure 6: Effect of Connecting Tubing on System

b. Measuring System Band-spreading Volume and
System Variance

This test should be performed on an HPLC system with a single
wavelength UV detector (not a Photodiode Array (PDA)).

1. Disconnect column from system and replace with a zero dead

volume union.

2. Set flow rate to 1 mL/min.

3. Dilute a test mix in mobile phase to give a detector sensitivity of

0.5 - 1.0 AUFS (system start up test mix can be used which contains
uracil, ethyl and propyl parabens; Waters part number WAT034544).

4. Inject 2 to 5 µL of this solution.

SLIPFREE

®

Type

Tubing Internal Diameter

Tubing Length

0.005”

0.010”

0.020”

Single 6 cm

PSL 618000

PSL 618006

PSL 618012

Single 10 cm

PSL 618002

PSL 618008

PSL 618014

Single 20 cm

PSL 618004

PSL 618010

PSL 618016

Double 6 cm

PSL 618001

PSL 618007

PSL 618013

Double 10 cm

PSL 618003

PSL 618009

PSL 618015

Double 20 cm

PSL 618005

PSL 618001

PSL 618017

Diluted/Distorted Sample Band

0.005 inches

0.020 inches

0.040 inches