beautypg.com

Waters CORTECS 1.6 um Columns User Manual

Page 9

background image

Waters Corporation
34 Maple Street
Milford, MA 01757 U.S.A.
T: 1 508 478 2000
F: 1 508 872 1990
www.waters.com

©2013 Waters Corporation. Waters, ACQUITY UPLC, UPLC, Oasis and
SepPak are registered trademarks of Waters Corporation. CORTECS,
VanGuard, eCord and The Science of What’s Possible are trademarks
of Waters Corporation. All other trademarks are the property of their
respective owners.

June 2013 720004750EN Rev. A VW-PDF

Mobile-Phase Considerations
1. Always maintain at least 3% polar solvent in the mobile

phase or gradient (e.g., 3% aqueous/3% methanol or 2%
aqueous/1% methanol, etc.). This ensures that the CORTECS
particle is always hydrated.

2. Maintain at least 40% organic solvent (e.g., acetonitrile) in

your mobile phase or gradient.

3. Avoid phosphate salt buffers to avoid precipitation in HILIC

mobile phases. Phosphoric acid is okay.

4. Buffers such as ammonium formate or ammonium acetate

will produce more reproducible results than additives such
as formic acid or acetic acid. If an additive (e.g., formic acid,
etc.) must be used instead of a buffer, use 0.2% (v:v) instead
of 0.1%.

5. For best peak shape, maintain a buffer concentration of 10 mM

in your mobile phase/gradient at all times.

Injection solvents
1. If possible, injection solvents should be 95% acetonitrile.

The polar solvent (i.e., water, methanol, isopropanol) should
be minimized to 25% of the total volume.

2. A generic injection solvent is 75:25 acetonitrile/methanol.

This is a good compromise between analyte solubility and
peak shape.

3. Avoid water and dimethylsulfoxide (DMSO) as injection sol-

vents. These solvents will produce very poor peak shapes.

4. Exchange water or DMSO with acetonitrile by using

reversed-phase solid-phase extraction (SPE). If this is not
possible, dilute the water or DMSO with organic solvent.

Miscellaneous tips
1. CORTECS HILIC Columns are designed to retain very polar

bases. Acidic, neutral and/or non-polar compounds will have
limited retention.

2. Optimal flow rates for small (<200 daltons) very polar bases

are in the 0.4 to 0.8 mL/min range with the CORTECS HILIC
Columns.

3. In HILIC, it is important to remember that water is the strongest

solvent. Therefore, it must be eliminated or minimized in the
injection solvent.

4. For initial scouting conditions, run a gradient from

95% acetonitrile to 50% acetonitrile. If no retention occurs, run
isocratically with 95:3:2 acetonitrile/methanol/aqueous buffer.

5. Alternate polar solvents such as methanol, acetone or isopropanol

can also be used in place of water to increase retention.

6. If using an ACQUITY UPLC System, the weak needle wash

should closely match the % organic present in the initial
mobile-phase conditions, otherwise, analyte peak shape or
retention may suffer.