beautypg.com

Care and use manual – Waters Dye QCRM Kit User Manual

Page 5

background image

Dye Quality Control Reference Material Kit

5

[ CARE AND USE MANUAL ]

• For a UV-directed system, enter 418 in the Wavelength A

column and 590 in the Wavelength B column. In the Fraction
Trigger 1 column, enter Wavelength A for a PDA detector
or UV1 for a UV detector, such as the Waters 2489 dual
wavelength absorbance detector. In the Fraction Trigger 2
column, enter Wavelength B for a PDA detector, or UV2 for
an UV detector.

Tip: The Mass and Wavelength columns are for information only,
unless the corresponding fraction information is entered in the
Fraction Trigger column.

6. For systems that include a ZMD mass detector, set the

instrument’s multiplier voltage to 550 V. For systems that
include a ZQ

mass detector, set the multiplier voltage to

500 V. For systems that include a 3100 or SQD2 mass
detector, retain the default gain setting of 0.1.

7. From the shortcut bar, click FractionLynx > Collection

Control, and verify that your system is activated.

To set the FractionLynx method parameters for a system test

1. In the FractionLynx main window, click Edit FractionLynx method.

2. In the FractionLynx method, click each tab to set the parameters

you want to define for this fraction collection method.

Exception: The FractionLynx method parameters for the UV-based
system test apply to Waters UV detectors. When using the detectors
of other manufacturers, the threshold values can differ slightly. They
might also differ from system to system. Use the values specified in
the Table 3 as guidelines.

Table 3. FractionLynx Method Settings

Parameter Tab

Settings You Can Specify

General

Set Fraction Collection to On and Peak Type to
Preparative

Select the Max Fractions per injection check box,
and specify a value of 10

Select the Max Tubes per injection check box, and
specify a value of 10

Rinse time = 0

Span = 0.5 amu

Timing

Solvent Front Delay = 0 seconds

Split/Collector Delay = seconds, where x = the
split collector delay determined earlier

Volume

Minimum Fraction Width = 3 seconds

Maximum Fraction Width = 120 seconds

Maximum Tube Fill = 90 percent

3. Based on your detector, enter the values specified in the

table below.

Tip: Where values are not specified in Table 4, retain FractionLynx
method’s defaults.


Table 4. FractionLynx Method Settings Based on Detector Type

Detector
Type

FractionLynx
Method Tab

Settings You Can Specify

2489

UV

MIT = 1000

Peak Start = Leading Edge Gradient %
with a value of 30

Terminate Peak = Below Gradient %
with a value of 60

PDA

PDA

Span = 3 nm

MIT = 5,000

Peak Start = Leading Edge Gradient %
with a value of 30

Terminate Peak = Below Gradient %
with a value of 60

Mass Spec

ES+

ES+ Ion Adducts = 1

MIT = 5,000,000

Peak Start = Use MIT only

Terminate Peak = Use MIT only

Analog

Analog

MIT = 5,000

Peak Start = Leading Edge Gradient %
with a value of 30

Terminate Peak = Below Gradient %
with a value of 60