beautypg.com

E. measuring system dwell volume, F. column equilibration – Waters CORTECS 2.7 um Columns User Manual

Page 3

background image

3

CORTECS 2.7 µm Columns

Table 1. Expected System Band Spread Volumes for Waters
LC Systems

Recommended

CORTECS Column i.d.

System

Band

Spread

1st

Choice

2nd

Choice

Alliance

®

2695 HPLC

29 µL

4.6 mm

3.0 mm

ACQUITY UPLC

12 µL

2.1 mm

3.0 mm

ACQUITY UPLC H-Class

9 µL

2.1 mm

3.0 mm

ACQUITY UPLC I-Class (FTN)

7.5 µL

2.1 mm

3.0 mm

ACQUITY UPLC I-Class (FL)

5.5 µL

2.1 mm

1.0 mm

Non-Waters UHPLC System

15–25 µL

3.0 mm

2.1 mm

e. Measuring System Dwell Volume

Dwell volume is different than system band spreading. System
dwell volume is a measurement of the volume it takes for the
initial gradient conditions to reach the head of the column. This
value is necessary when transferring a method between different
chromatographic systems.

1. Disconnect the column from the system and replace with a zero

dead volume union.

2. Use acetonitrile as mobile phase A, and acetonitrile containing

0.05 mg/mL uracil as mobile phase B.

3. Use the flow rate in the original method and the intended flow

rate on the target instrument.

4. Collect detector baseline using 100% mobile phase A for

5 minutes.

5. At 5.00 minutes, program a step to 100% mobile phase B,

and collect data for an additional 5 minutes.

6. Measure absorbance difference between 100% A and 100% B.

7. Measure time at 50% of that absorbance difference.

8. Calculate time difference between start of step and 50% point.

9. Multiply time difference by the flow rate to calculate

system volume.









5.69 minutes

- 5.00 minutes

0.69 minutes

50%

Time = 5.69 minutes

0.00

0.05

0.10

0.15

0.20

0.25

0.30

0.35

0.40

0.45

0.50

0.55

0.60

0.65

0.70

0

2

4

6

8

10

12

14

16

18

20 min

Programmed time = 5.00 minutes

System Volume:

0.69 min x 1.5 mL/min = 1.04 mL

Flow Rate = 1.5 mL/min

AU

Figure 3. Measuring System Dwell Volume.

Table 2. Expected System Dwell Volumes for Waters Systems

System

Dwell Volume

Alliance 2695 HPLC

900 µL

ACQUITY UPLC

120 µL

ACQUITY UPLC H-Class

350 µL

ACQUITY UPLC I-Class (FTN)

100 µL

ACQUITY UPLC I-Class (FL)

95 µL

f. Column Equilibration

CORTECS 2.7 µm Columns are shipped in 100% acetonitrile. It is
important to ensure mobile phase compatibility before changing
to a different mobile phase system. To avoid precipitating mobile
phase buffers within the column or system, flush the column with
five column volumes of a waters/organic solvent mixture, using
the same, or lower, solvent content as in the desired buffered
mobile phase (for example, flush the column and system with 60%
methanol in water prior to introducing 60% methanol/40% buffer
mobile phase).

Note. If mobile phase additives (i.e., ion-pairing reagents) are
present in low concentrations (<0.2% v/v), 100 to 200 column
volumes may be required for complete equilibration. In addition,
mobile phases that contain formate (i.e., ammonium formate,
formic acid) may require extended equilibration times.

For reversed-phase separations, equilibrate the column with a
minimum of 10 column volumes of the mobile phase to be used
(refer to Table 3 for a list of column volumes). The column may be
considered fully equilibrated once a constant backpressure and a
stable detector baseline is achieved.