Care and use manual, V. oasis sorbent selection and p rotocol chart, Select one of the four oasis sorbents – Waters Oasis Mixed-Mode Ion-Exchange Cartridges and 96-Well Plates User Manual

Page 4: Apply the indicated protocol [1 or 2, Oasis 2x4 strategy

background image

Table 1: Recommended Volume for Generic Methods (assuming 1:1 dilution)

Cartridges

96-well plate

µElution Plate

Cartridges size/
Sorbent mass

1 cc

3 cc

6 cc 12 cc 20 cc

35 cc

5 mg

10 mg

30 mg

60 mg

2 mg

Condition/
Equilibration (ml)

1 ml

2 ml

3 ml

5 ml 10 ml

50 ml

0.2 ml

0.5 ml

0.5 -1.0 ml

0.5 -2.0 ml

0.2 ml

Maximum load of
matrix and dilution

1 ml

2 ml

5 ml 15 ml 30 ml 100 ml

0.5 ml

1 ml

1-2 ml

1-2 ml

0.025-0.75 ml

Wash (ml)

1 ml

2 ml

4 ml

5 ml 10 ml

40 ml

0.2 ml

0.5 ml

0.5 -1.0 ml

1 - 2 ml

0.2 ml

Elute (ml)

1 ml

2 ml

4 ml

5 ml 10 ml

60 ml

0.05-0.2 ml 0.15-0.3 ml

0.4-1.0 ml

0.8-2.0 ml

0.025-0.10 ml

Note: Above listed sample volumes are for biological samples; for certain types of samples (i.e. drinking water) up to 20 times of sample solution is possible.
Note: Solution should be made fresh daily.

Table 2: Load Volumes for Large Volume WATER Analysis

Cartridge size/Sorbent mass

1 cc

3 cc

6cc (200 mg, 30 µm)

6cc (200 mg, 60 µm)

12 cc

20 cc

35 cc

Load (ml water*) (total of
matrix and dilution)

50 ml

200 ml

500 ml

1000 ml

1000 ml

2000 ml

5000 ml

V. OASIS SORBENT SELECTION AND P ROTOCOL CHART

The Oasis 2x4 Method is a simple, logical approach to the
selection of an SPE sorbent and protocol. Two protocols and
four sorbents provide the flexibility to extract acids, bases,
and neutrals with high SPE recoveries while removing matrix
components that may interfere with analysis.

Follow the simple steps outlined in this flow chart to achieve
high recoveries and the cleanest extracts:

Characterize your analyte [Neutral, Acid or Base, pK

a

]

Select one of the four Oasis Sorbents

Apply the indicated Protocol [1 or 2]

Determine SPE recoveries by LC analysis


*Recoveries for all acids, bases and neutral compounds
are greater than 85% with RSD’s less than 5%.

Strong

Acids

Weak

Acids

Weak

Bases

Strong

Bases

Weaker

Acids

Neutrals

Weaker

Bases

Oasis 2x4 Strategy

Only 2 protocols and 4 sorbents to

analyze all types of compounds:

acids, bases, and neutrals

For Weak Bases

pKa 2–10:

Use

Oasis MCX

For Weak Acids

pKa 2–8:

Use

Oasis MAX

For Strong Acids

pKa <1:

Use

Oasis WAX

For Strong Bases

pKa >10:

Use

Oasis WCX

Prepare Sample

Prepare Sample

Condition, Equilibrate,

and Load Sample

Condition, Equilibrate,

and Load Sample

Wash:

2% HCOOH

Wash:

5% NH

4

OH

Elute 1:

100% CH

3

OH

Elute 1:

100% CH

3

OH

Elute 2: 5% NH

4

OH

in CH

3

OH

Elute 2: 2% HCOOH

in CH

3

OH

Apply Protocol 1

Apply Protocol 2

4

[ CARE AND USE MANUAL ]

Oasis Mixed-Mode Ion-Exchange Cartridges and 96-Well Plates