filter that is already installed in the gas line. The gas flow rate is
approximately 20 liters/minute.
4. Install the oxygen cylinder in the same manner as outlined for the nitrogen
cylinder installation in Section 8.7.1, using the incubator’s O
2
inlet, of
course. For this application, you will need another two-stage pressure
regulator and another inline pressure regulator.
NOTE:
If the programmed O
2
level is close to the ambient oxygen, it may be
necessary to reduce the cylinder pressure below 1 bar to stop the oxygen
level from undershooting the programmed value.
Be sure to humidify the incubator and leave it overnight to stabilize before
proceeding further.
To automatically calibrate the sensor to atmospheric oxygen levels, select
OXYGEN SENSOR – REF TO ATMOSPHERE
in the
USER
menu, and then follow
the onscreen instructions.
The oxygen reading is automatically adjusted to 19.7%, which is the true
reading taking into account the Relative Humidity in the chamber.
8.7.3 Enabling or Disabling O
2
Control
To enable oxygen control:
1. Select the
USER
screen by pressing the
USER
function key.
2. Use the direction keys to select
MANUAL DISABLE
.
3. Press the
ENTER
function key, then use the direction keys to select
ON
for
Oxygen Control.
4. Press the
ENTER
function key again.
Program the required oxygen level in the
PROG
screen, following the onscreen
instructions.
To disable oxygen control, follow the pattern used to enable it, but this time
select
USER
,
MANUAL DISABLE
,
ENTER
,
OFF
and then
ENTER
again.