New Brunswick Scientific

User’s Guide

filter that is already installed in the gas line. The gas flow rate is
approximately 20 liters/minute.

4. Install the oxygen cylinder in the same manner as outlined for the nitrogen

cylinder installation in Section 8.7.1, using the incubator’s O

2

inlet, of

course. For this application, you will need another two-stage pressure
regulator and another inline pressure regulator.

NOTE:

If the programmed O

2

level is close to the ambient oxygen, it may be

necessary to reduce the cylinder pressure below 1 bar to stop the oxygen
level from undershooting the programmed value.

Be sure to humidify the incubator and leave it overnight to stabilize before
proceeding further.

To automatically calibrate the sensor to atmospheric oxygen levels, select

OXYGEN SENSOR – REF TO ATMOSPHERE

in the

USER

menu, and then follow

the onscreen instructions.

The oxygen reading is automatically adjusted to 19.7%, which is the true
reading taking into account the Relative Humidity in the chamber.


8.7.3 Enabling or Disabling O

2

Control

To enable oxygen control:
1. Select the

USER

screen by pressing the

USER

function key.

2. Use the direction keys to select

MANUAL DISABLE

.

3. Press the

ENTER

function key, then use the direction keys to select

ON

for

Oxygen Control.

4. Press the

ENTER

function key again.


Program the required oxygen level in the

PROG

screen, following the onscreen

instructions.

To disable oxygen control, follow the pattern used to enable it, but this time
select

USER

,

MANUAL DISABLE

,

ENTER

,

OFF

and then

ENTER

again.