New Brunswick Scientific

User’s Guide

ƒ

If chamber conditions recover within the programmed alarm delay time, the

Alarm System is immediately re-armed. After the delay, the Alarm System
is armed and if the temperature and CO

2

are outside the Alarm High & Low

settings, the alarm will be activated.


ƒ

If an alarm occurs and the chamber subsequently recovers, the alarm stops

and the system is re-armed. Details of the Alarm Event are stored in the
Datalogger.

Figure 13: Chamber Alarm System Flow Chart

15 MINUTES

ALARM DELAY

TIMEOUT

COMPLETED

ARE TEMPERATURES AND CO2

LEVELS WITHIN +/- 0.1 OF SET

POINT?

CHANGE

PROGRAMMED

VALUES

OPEN INNER DOOR

DISARM ALARMS

CLOSE INNER

DOOR

15 MINUTES

ALARM DELAY

TIMEOUT STARTED

IS TEMPERATURE OR CO2 LEVEL

MORE THAN +/- 0.5 OUTSIDE

PROGRAMMED POINTS?

CANCEL ALARM BY

PRESSING ANY KEY

ALARMS

CANCELLED

VISUAL AND AUDIBLE

ALARMS CANCELLED,

MESSAGE LEFT ON SCREEN

CANCEL

MESSAGE BY

PRESSING ANY

KEY

NO

YES

YES

SWITCH ON

ARM ALARMS

VISIBLE AND AUDIBLE

ALARM - MESSAGE ON

SCREEN

HAVE CONDITIONS

RECOVERED TO

WITHIN +/- 0.1?

YES

NO

NO