Warning – Eppendorf R Series 48R CO2 Incubator User Manual

Page 14

background image

14

CO48R-0050

Š Galaxy R Series 48 R Personal Size CO

2

Incubator

User’s Guide

Table 1: Accessories Provided

Quantity

Item

Notes

3

Non-tip Shelves

Packed Separately

2

Wire Shelf Racks

Packed Separately

1

Humidity Tray

Packed Separately

14

Silicone Rubber Suction Feet

Installed

1

White porous CO

2

Sensor Cover

Installed

1

Black Sensor Cover*

Installed

1

Power Cord

Packed in accessories bag

3 meters,

9.8 feet

PVC Tubing, ~1/4-inch or 6mm bore,
with an inline CO

2

HEPA-filter

connected, ready for use

Packed in accessories bag

2

Hose Clips

Packed in accessories bag

1

AutoZero HEPA Filter

Packed in accessories bag

4

Adjustable Feet

Packed in accessories bag

4

Anti-slip Pads for adjustable feet

Packed in accessories bag

1 User

Manual

Provided

* There is a holder at the rear of the incubator to store the black sensor cover.

WARNING!

Anytime you touch or handle the white CO

2

sensor cover, be sure to

wear gloves, and do not later touch those gloves to your face. Discard

or wash the gloves.

2.3 Utilities

In order to use the incubator, you will need:

Table 2: Utilities

Utility

Requirement

100/120V, 50/60 Hz grounded electrical supply with minimum capacity of 6
amps (or 8 amps for High Temperature Disinfection Models)

OR

Electricity

220/240V, 50 Hz grounded electrical supply with minimum capacity of 3 amps
(or 5 amps for High Temperature Disinfection Models)

CO

2

Gas

Cylinder with 100% CO

2

vapor withdrawal, together with a two-stage regulator

for pressure control to 5 psi or 0.35 bar

2.4 Location

The incubator is designed to operate at a chamber temperature of 4.0°C above ambient,
and at an absolute minimum ambient temperature of 15°C if the incubator is being used
at 37°C. Care should be taken to avoid placing the incubator in a position that may
affect its performance, such as those listed below.