Warning – Eppendorf R Series 48R CO2 Incubator User Manual
Page 14
14
CO48R-0050
Galaxy R Series 48 R Personal Size CO
2
Incubator
User’s Guide
Table 1: Accessories Provided
Quantity
Item
Notes
3
Non-tip Shelves
Packed Separately
2
Wire Shelf Racks
Packed Separately
1
Humidity Tray
Packed Separately
14
Silicone Rubber Suction Feet
Installed
1
White porous CO
2
Sensor Cover
Installed
1
Black Sensor Cover*
Installed
1
Power Cord
Packed in accessories bag
3 meters,
9.8 feet
PVC Tubing, ~1/4-inch or 6mm bore,
with an inline CO
2
HEPA-filter
connected, ready for use
Packed in accessories bag
2
Hose Clips
Packed in accessories bag
1
AutoZero HEPA Filter
Packed in accessories bag
4
Adjustable Feet
Packed in accessories bag
4
Anti-slip Pads for adjustable feet
Packed in accessories bag
1 User
Manual
Provided
* There is a holder at the rear of the incubator to store the black sensor cover.
WARNING!
Anytime you touch or handle the white CO
2
sensor cover, be sure to
wear gloves, and do not later touch those gloves to your face. Discard
or wash the gloves.
2.3 Utilities
In order to use the incubator, you will need:
Table 2: Utilities
Utility
Requirement
100/120V, 50/60 Hz grounded electrical supply with minimum capacity of 6
amps (or 8 amps for High Temperature Disinfection Models)
OR
Electricity
220/240V, 50 Hz grounded electrical supply with minimum capacity of 3 amps
(or 5 amps for High Temperature Disinfection Models)
CO
2
Gas
Cylinder with 100% CO
2
vapor withdrawal, together with a two-stage regulator
for pressure control to 5 psi or 0.35 bar
2.4 Location
The incubator is designed to operate at a chamber temperature of 4.0°C above ambient,
and at an absolute minimum ambient temperature of 15°C if the incubator is being used
at 37°C. Care should be taken to avoid placing the incubator in a position that may
affect its performance, such as those listed below.