8.1 Options
Some option combinations are not possible, others may incur extra cost. Please inquire
before ordering.
Options
NBS Part Number
Field
Retrofitable?
High Temperature Disinfection
CO170R-1000
No
O
2
Control, 1-19%
CO170R-0200
No
O
2
Control, 0.1-19%
P0628-5410
No
O
2
Control, 1-195%
P0628-5400
No
Cooling System
P0628-6810
No
4 Split Inner Glass Door
P0628-6780
Yes
8 Split Inner Glass Door
P0628-6781
Yes
Copper Chamber
P0628-5612
No
Humidity Package
P0628-6820
No
BMS Relay
P0628-5540
Yes
IP66 Power Receptacle
P0628-5612
No
8.2 High
Temperature Disinfection
The High Temperature Disinfection option is designed to heat the internal chamber to
120°C, maintain that temperature for 4 hours, and then allow the chamber to cool down
to 37°C or the programmed temperature (if different from 37°C) when normal control
takes over. The cycle is designed to disinfect all internal surfaces and components with
the exception of the Oxygen control sensor where supplied.
1. The incubator should be cleaned, disinfectd chemically and dried thoroughly before
starting the cycle (follow the instructions for Cleaning in Section 5).
2. The black protective cover must be removed before starting a cycle (the white
porous cover can remain in place). The shelves, shelf racks, humidity tray and
silicone rubber suction cups should all be in place during the cycle. If the Humidity
Tray Warning Option is fitted, ensure the sensor is placed on the bottom shelf prior
to starting the cycle. This is to protect the sensor from the direct heat at the bottom
of the incubator. The incubator MUST be clean and dry, and the humidity tray
MUST be empty, clean and dry, to continue. If it is not, permanent damage can
occur to the incubator that may void your warranty.