beautypg.com

Eppendorf R Series 48R CO2 Incubator User Manual

Page 24

background image

24

CO48R-0050

Š Galaxy R Series 48 R Personal Size CO

2

Incubator

User’s Guide

3.6.1

DATE/TIME ADJUST

The date and time are factory-set and will only require adjustment if you are in
a different time zone, or when you change your clocks to Daylight Saving
Time and back again to Standard time. You may also select the style of
display for the date.

A battery back-up system ensures that the correct time/date setting is never lost.


3.6.2

AUDIO

VOLUME

ADJUST

The audible volume can be adjusted to your own preferences.



3.6.3

PROGRAMMABLE

CO

2

AUTOZERO

When you select this feature, the

PROGRAM CO

2

AUTOZERO

screen (see Figure

7 below) allows you to program the AutoZero frequency and time, or run the
AutoZero function manually.

The AutoZero System automatically re-references the CO

2

Sensor to

atmospheric CO

2

in the following way:

1. A pump activates for two minutes, pumping HEPA-filtered atmosphere at

0.3 liters/minute into the sensor’s measuring chamber. This displaces the
chamber atmosphere completely from the sensor.

Figure 7: PROGRAM CO

2

AUTOZERO Screen

NOTE:

This procedure does not affect the internal chamber environment. It will
not affect your cell culture as the procedure is being performed.