Eppendorf R Series 48R CO2 Incubator User Manual
Page 26

26
CO48R-0050
Galaxy R Series 48 R Personal Size CO
2
Incubator
User’s Guide
3.6.7
DISINFECTION
(Model Option)
If the incubator is supplied with the High Temperature Disinfection option, the
menu item
DISINFECTION
will be displayed. This feature activates the
disinfection cycle of the incubator. The disinfection cycle heats the inner
chamber to 120
o
C, holds that temperature for 4 hours, then cools the chamber to
the selected temperature setpoint. All of the interior components (with the
exception of certain sensors) can be left in place during the cycle to ensure that
everything within the chamber is disinfectd prior to resumption of activity. For
a full explanation of this feature, see section 10.1.
3.7
DATALOGGER
Screen
The
DATALOGGER
Screen displays the following information:
Figure 8: DATALOGGER Screen
This section explains these features.
3.7.1
ALARM
EVENTS
The following alarm events are recorded in the order in which they occurred,
with the most recent event displayed at the top.
Power ON / OFF
Chamber Temperature High / Low (programmed value)
CO
2
Level High / Low (programmed value)
CO
2
Supply Failure
All System Alarms
CO
2
AutoZero (A/Z) Adjustments
Oxygen and Relative Humidity (R/H) Alarms (where these options are
installed)