DISINFECTION

(Model Option)

If the incubator is supplied with the High Temperature Disinfection option, the
menu item

DISINFECTION

will be displayed. This feature activates the

disinfection cycle of the incubator. The disinfection cycle heats the inner
chamber to 120

o

C, holds that temperature for 4 hours, then cools the chamber to

the selected temperature setpoint. All of the interior components (with the
exception of certain sensors) can be left in place during the cycle to ensure that
everything within the chamber is disinfectd prior to resumption of activity. For
a full explanation of this feature, see section 10.1.

3.7

DATALOGGER

Screen

The

DATALOGGER

Screen displays the following information:

Figure 8: DATALOGGER Screen

This section explains these features.


3.7.1

ALARM

EVENTS

The following alarm events are recorded in the order in which they occurred,
with the most recent event displayed at the top.

ƒ

Power ON / OFF

ƒ

Chamber Temperature High / Low (programmed value)

ƒ

CO

2

Level High / Low (programmed value)

ƒ

CO

2

Supply Failure

ƒ

All System Alarms

ƒ

CO

2

AutoZero (A/Z) Adjustments

ƒ

Oxygen and Relative Humidity (R/H) Alarms (where these options are

installed)