2. If required, you can take a sample of the gas inside the chamber using the CO

2

sample port, and check it using a CO

2

gas analyzer. The CO

2

sample port is located

at the top of the rear panel.

If you conduct a sampling, please do the following:

ƒ

Turn off the CO

2

gas by re-programming the setpoint for CO

2

to 0.0% to prevent

CO

2

from being injected into the chamber and giving a false reading.

ƒ

Use a flow rate that is

≤ 0.5 liters/minute to take a sample.

ƒ

Keep the door closed to minimize heat loss.

ƒ

Remember to reset the CO

2

setpoint to the desired level after sampling.

We recommend that you perform a CO

2

AutoZero prior to sampling (see Section

3.5).

3. We also recommend that you AutoZero the CO

2

system at least once every 28 days

to ensure that CO

2

level is correct.