Eppendorf R Series 48R CO2 Incubator User Manual
Page 57

57
New Brunswick Scientific
User’s Guide
1. Remove the black plastic protective cover (making sure that the
hydrophobic filter cap is not removed with it) from the port inside the
chamber. Retain the cover for use when you clean the chamber.
2. Open the nitrogen gas supply from the tank and set the N
2
tank’s outlet
pressure gauge to 1.5 bar.
3. Set the nitrogen regulator to 1 bar (14.5 psi). The nitrogen will be fed into
the incubator through a HEPA filter that is already installed in the gas line.
The gas flow rate is approximately 20 liters/minute.
NOTE:
If the programmed O
2
level is close to the ambient oxygen, it may be
necessary to reduce the cylinder pressure below 1 bar to stop the oxygen
level from undershooting the programmed value.
NOTE:
When working at 0.1% O
2
, set the CO
2
pressure to 0.2 bar (2.9 psi).
Be sure to humidify the incubator and leave it overnight to stabilize before
proceeding further.
To automatically calibrate the sensor to atmospheric oxygen levels, select
OXYGEN SENSOR – REF TO ATMOSPHERE
in the
USER
menu, and then follow
the onscreen instructions.
The oxygen reading is automatically adjusted to 19.7%, which is the true
reading taking into account the Relative Humidity in the chamber.
8.6.3 Enabling or Disabling O
2
Control
To enable oxygen control:
1. Select the
USER
screen by pressing the
USER
function key.
2. Use the direction keys to select
MANUAL DISABLE
.
3. Press the
ENTER
function key, then use the direction keys to select
ON
for
Oxygen Control.
4. Press the
ENTER
function key again.
Program the required oxygen level in the
PROG
screen, following the onscreen
instructions.
To disable oxygen control, follow the pattern used to enable it, but this time
select
USER
,
MANUAL DISABLE
,
ENTER
,
OFF
and then
ENTER
again.