1. Remove the black plastic protective cover (making sure that the

hydrophobic filter cap is not removed with it) from the port inside the
chamber. Retain the cover for use when you clean the chamber.

2. Open the nitrogen gas supply from the tank and set the N

2

tank’s outlet

pressure gauge to 1.5 bar.

3. Set the nitrogen regulator to 1 bar (14.5 psi). The nitrogen will be fed into

the incubator through a HEPA filter that is already installed in the gas line.
The gas flow rate is approximately 20 liters/minute.

NOTE:

If the programmed O

2

level is close to the ambient oxygen, it may be

necessary to reduce the cylinder pressure below 1 bar to stop the oxygen
level from undershooting the programmed value.

NOTE:

When working at 0.1% O

2

, set the CO

2

pressure to 0.2 bar (2.9 psi).

Be sure to humidify the incubator and leave it overnight to stabilize before
proceeding further.

To automatically calibrate the sensor to atmospheric oxygen levels, select

OXYGEN SENSOR – REF TO ATMOSPHERE

in the

USER

menu, and then follow

the onscreen instructions.

The oxygen reading is automatically adjusted to 19.7%, which is the true
reading taking into account the Relative Humidity in the chamber.

8.6.3 Enabling or Disabling O

2

Control

To enable oxygen control:
1. Select the

USER

screen by pressing the

USER

function key.

2. Use the direction keys to select

MANUAL DISABLE

.

3. Press the

ENTER

function key, then use the direction keys to select

ON

for

Oxygen Control.

4. Press the

ENTER

function key again.

Program the required oxygen level in the

PROG

screen, following the onscreen

instructions.

To disable oxygen control, follow the pattern used to enable it, but this time
select

USER

,

MANUAL DISABLE

,

ENTER

,

OFF

and then

ENTER

again.