Eppendorf R Series 48R CO2 Incubator User Manual
Page 22
22
CO48R-0050
Galaxy R Series 48 R Personal Size CO
2
Incubator
User’s Guide
NOTE:
If a temperature
>1°C below the actual chamber temperature is set, the
independent over-temperature system will activate. This is indicated by a
warning message on the display. When the chamber cools down, normal
operation will resume. Programming a higher temperature can also
deactivate the warning.
3.4.2 User Access Code
A
USER ACCESS CODE
is programmable, if required. The User Access Code
allows you to restrict access to the
PROG
,
USER
, and
ALARM
screens (where
settings can be changed) to authorized persons only.
To set the User Access Code:
1. In the
PROG
screen (accessed by pressing the
PROG
function key), the user
access code will be displayed as a series of four asterisks.
2. Use the left and right direction keys to move to each code position, and the
up and down direction keys to select a number from 0 to 9.
3. Once the number is selected, press the
ENTER
function key to save the code.
4. After returning to the main screen, programming access will require the
code to make any further programming changes.
To cancel the access code:
1. In the
PROG
screen, enter the current access code.
2. Now program 0000 as the new access code.
3. Press the
ENTER
function key to save the change.
4. The code is now cancelled and programming is no longer restricted.
NOTE:
If the access code has been misplaced, you will be unable to make
changes to your incubator’s settings. Contact customer service or your
service representative for instructions on how to regain access to your
incubator.