8.6.4 Alarms

The alarm levels are set automatically to ± 0.5% above and below the
programmed value, but these points can be altered in the

ALARM

screen. In

addition, the rearming of the alarm can be delayed until the Programmed Value
is achieved by selecting the relevant option in the

ALARM

screen.

For example, it is possible at low oxygen levels that the CO

2

and O

2

levels

might not fully recover within the Alarm Limits after the 15-minute

DELAY IN

ARMING AFTER DOOR OPENING

period has elapsed, so this time period can be

increased in the

ALARM

screen to suit individual circumstances.


8.6.5 Referencing to Atmosphere

To reference the oxygen sensor to atmospheric oxygen levels, see Section 8.5.4.


8.6.6 Programming Desired O

2

Level

Program the required oxygen level in the

PROG

screen, following the onscreen

instructions.

If you are running an O

2

level programmed between 0.1-0.9%,

you should know that the control system is set to operate in the following way to
minimize N

2

consumption after the glass door has been opened:


• The N

2

valve is switched on continuously until the O

2

level is within 0.1%

of setpoint.

• The CO

2

valve is then switched on to allow the CO

2

level to reach setpoint.

If the O

2

level is above setpoint 15 minutes after the N

2

valve has been

switched off, it is switched back on for 40 seconds and the CO

2

valve is

switched on for 20 seconds. The CO

2

valve will then pulse until setpoint is

reached.

• The process described above will repeat itself until the O

2

setpoint is

reached.

• The same process will also repeat if the O

2

level rises above setpoint, and if

the O

2

level should rise toward 0.2% above setpoint, the N

2

valve will open

again continuously until the O

2

level returns to setpoint.

• The CO

2

AutoZero, which would normally take place after a CO

2

alarm, will

be cancelled to avoid the introduction of additional O

2

into the chamber.

For the same reason, we recommend canceling the programmed CO

2

AutoZero.