Eppendorf R Series 48R CO2 Incubator User Manual
Page 44

44
CO48R-0050
Galaxy R Series 48 R Personal Size CO
2
Incubator
User’s Guide
• failed for any reason,
DISINFECTION FAILED [CODE: XX]
. The codes are shown
in Table 4a and described in Table 4b. If this happens, please note the failure
code and contact your service representative for advice.
Table 4a: Disinfection Failure Codes & Descriptions
Failure Code
Failure Code Description
(see Table 4b below)
01
Z
02
W
03
W, Z
04
X
05
X, Z
06
W, X
07
W, X, Z
08
Y
09
Y, Z
0A
W, Y
0B
W, Y, Z
0C
X, Y
0D
X, Y, Z
0E
W, X, Y
0F
W, X, Y, Z
Table 4b: Disinfection Failure Code Explanations
Failure Code
Description
Explanation
W
Temperature drop during warm-up period: indicates
the temperature fell more than 2
°C during the
heating phase over a 60-second period.
X
Temperature drop during 4-hour period: i
ndicates
the temperature fell below 118.0°C during the
disinfection phase.
Y
Temperature increase during cool-down phase:
i
ndicates the temperature rose by more than 2°C
during the cooling phase over a 60-second period.
Z
Cancel key pressed.
NOTE:
If the incubator power is cycled
OFF
then
ON
during a disinfection cycle
due to a power outage, the incubator will power up as normal. This
condition will be indicated by the absence of a completed disinfection
status message (
DISINFECTION COMPLETED OK
).
…continued…