3o o – Eppendorf R Series 48R CO2 Incubator User Manual
Page 19
19
New Brunswick Scientific
User’s Guide
3
3
O
O
P
P
E
E
R
R
A
A
T
T
I
I
O
O
N
N
3.1 Control
Panel
The control panel consists of an LCD display, five function keys & four direction keys
(see Figure 5):
Figure 5: Control Panel
1 Direction keys
2 Function keys
The purpose of each Function Key is identified at the bottom of the display (above each
key); the function may change from screen to screen.
The cluster of four arrowed Direction Keys will move the cursor around the screen and
adjust values.
The
HELP
file contains most of the information in this User Manual, together with more
detailed troubleshooting information.
3.2 Preparing for Operation
1. Remove the black protective cover from the CO
2
sensor (located on the back
surface of the internal chamber), and store it for use when you clean the incubator.
There is a storage holder for the cover on the back on the incubator.
2. Ensure that the white porous sensor cover remains in place.
3. Using the power cord provided, connect the incubator to a grounded power supply.
4. Switch the incubator ON using the on/off switch at the rear of the cabinet. The
display will illuminate immediately.
5. Turn on the CO
2
gas supply with the pressure regulator set to 5 psi or 0.35 bar.
2
1