Eppendorf R Series 48R CO2 Incubator User Manual

Page 29

background image

29

New Brunswick Scientific

User’s Guide

3.7.6

DIAGNOSTIC

DOOR

ELEMENT

GRAPH

This graph records door element temperature over time to assist troubleshooting.


3.7.7

RESTART

GRAPHIC

RECORD

This feature removes the current graph and begins a new one. The data cannot
be recovered once it is deleted.

3.8

CHAMBER ALARMS

Menu Screen

The

CHAMBER ALARMS

programming screen (see Figure 11 below) allows the

various alarm options to be selected and modified. Press the S or T direction
key to move around the options and the W or X direction key to adjust values.
The temperature and CO

2

High and Low Alarm setpoints automatically adjust to

within

± 0.5 of the temperature and CO

2

setpoints. The alarm setpoints can also

be manually adjusted.

Figure 11: CHAMBER ALARMS Programming Screen


To arm the Chamber Alarms after a selectable delay:
1. Choose the option

ARM ALARMS WHEN AT SETPOINT:

2. Select

NO

for both

TEMP

and

CO2

(as shown in Figure 11 above)

3. Choose the option

DELAY IN ARMING AFTER DOOR OPEN

: and select the

desired delay (15 minutes in the sample screen above) to allow for
temperature and CO

2

recovery after the door has been opened.


Alternatively, the alarm system can be set to rearm only after the original
temperature and CO

2

setpoints have been achieved:

1. Choose the option

ARM ALARMS WHEN AT SETPOINT:

2. Select

YES

for both

TEMP

and

CO2

.