Š Galaxy R Series 48 R Personal Size CO

2

Incubator

User’s Guide

3. When

YES

is selected for this function, the

DELAY IN ARMING AFTER DOOR

OPEN

is ignored.


A

DOOR OPEN ALARM

: can be adjusted, choosing from seven preset durations

(45 seconds in Figure 11) to warn of an improperly closed door.

The

AUDIBLE

and

VISUAL

alarms can be adjusted from

OFF

to

ON

(which means

the alarm will be on continuously until it is acknowledged) in seven preset time
increments.

In the

OFF

position, any Chamber Alarms that occur will be displayed on the

screen without flashing and with the audible alarm inhibited.

Section 3.8.1, Chamber Alarm System Function, and Figure 13 explain in detail
how the Chamber Alarms operate.

3.8.1 Chamber Alarm System Function

When the incubator is switched ON, or after the temperature and CO

2

levels

have been reprogrammed, the Alarm System is inactive until the setpoint values
(± 0.1) are achieved, after which the Alarm System is armed. As setpoint is
achieved, the CO

2

and temperature alarms are individually armed.


If temperature and/or CO

2

levels deviate more than the programmed setpoints,

the display flashes, the audible alarm sounds and a message appears on the
screen. You can acknowledge the alarm by pressing any key.

Figure 12: CHAMBER ALARM Message


After setpoints have been achieved for the first time, when the inner door is
opened, the Alarm System is disabled; on closing the door, if selected, a
programmable alarm delay starts: