beautypg.com

Eppendorf R Series 48R CO2 Incubator User Manual

Page 60

background image

60

CO48R-0050

Š Galaxy R Series 48 R Personal Size CO

2

Incubator

User’s Guide

8.7.1 Setting Up the N

2

Tank

Before you set up your oxygen control, be sure you have the proper equipment
for your nitrogen supply:

• 2 cylinders of nitrogen, regulation size W
• 1 two-stage pressure regulator
• 1 Inline pressure regulator
• 6mm PVC tubing
• tubing clips

Set up the nitrogen tanks as follows:

1. Inspect them to be sure there are no leaks or other damage.
2. Check that the two-stage pressure regulator valve and the inline pressure

regulator valve are closed by trying to turn their knobs in a counter-
clockwise direction; the knobs should be at the stop point.

3. Place the inline pressure regulator close to the incubator.
4. Securely attach the two-stage pressure regulator to the nitrogen tank’s outlet

and to the inline pressure regulator’s inlet with tubing, securing both ends
with clips.

5. Connect the inline pressure regulator’s outlet to the incubator’s N

2

inlet on

the right side of the equipment tray at the rear of the incubator (see Figure
7
); secure both ends with clips.

6. Proceed as indicated in Section 8.7.2.

NOTE:

To add an automatic gas cylinder changeover instrument: with the
incubator turned off, connect its inlet to the two-stage pressure
regulator’s outlet, and the changeover instrument’s outlet to the inline
pressure regulator’s inlet with tubing, securing the ends with tubing clips.

8.7.2 Setting Up O

2

Control

Oxygen control is tailored such that both the O

2

and the CO

2

levels are achieved

at approximately the same time, via control of the O

2

or N

2

valve’s duty cycle.


1. Remove the black plastic protective cover (making sure that the

hydrophobic filter cap is not removed with it) from the port inside the
chamber. Retain the cover for use when you clean the chamber.

2. Open the nitrogen gas supply from the tank and set the N

2

tank’s outlet

pressure gauge to 1.5 bar.

3. Set the nitrogen regulator to 1 bar (14.5 psi)—but be sure to read the

NOTE below. The nitrogen will be fed into the incubator through a HEPA