7 data bytes 7 ascii characters, 2 data bytes 16-bit value, 6 = forward – Apex Digital 5708009-C User Manual

Page 96: 7 = delivered or load, 8 = external (dc bias), 1 data byte 8-bit value, 4 data bytes 32-bit value

background image

4-50

5708009-C

Advanced Energy

®

130

read

mainframe

software

version

number

Requests the version number of the
mainframe software. The returning packet
contains 7 ASCII characters—a 7-digit
number. This command is used in
conjunction with CMD 198 to obtain the
version/revision number of the mainframe
software.

0

7 data bytes

7 ASCII

characters

140

report host

time-out

value

Requests the serial Host port time-out value
(002 to 500 representing 0.02 to 5.00 s).

0

2 data bytes
16-bit value

154

report

regulation

mode

Requests regulation mode (set with CMD 3).
Returning values:

• 6 = Forward
• 7 = Delivered or Load
• 8 = External (DC Bias)

1 data byte

8-bit value

155

read control

method

Requests control mode (set by CMD 14).
Returning values: 2 = host, 4 = analog

0

1 data byte

8-bit value

156

read active

target

Requests the number of the active target (set
by CMD 11)

0

1 data byte

8-bit value

157

read target

life

Requests the amount of life remaining in the
target you specify (set by CMD 12). This
command requires 1 data byte to specify the
number of the target you request (1 to 4).

Note: A decimal is implied—100 = 1 kWh,

1 data byte

8-bit value

4 data bytes
32-bit value

Table 4-6. Host Port Commands for RS-232 with AE Bus (Continued)

Command

Description

Number of

Host Data

Bytes

Number of

Response

Data Bytes