Error messages, 14 information on ph measurement & electrode 53, 12 error messages – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 55

background image

51

12 ERROR

MESSAGES

The following table provides a guideline to enable diagnosis of possible problems indicated by

the messages generated by the CyberScan meter. The table also provides possible solutions

to the problems encountered.

ERROR MESSAGE

INDICATES

POSSIBLE CAUSE

CORRECTIVE
ACTION

Err. 1 (in primary
display)

Memory write error

Instrument too old (> 10
years) or hardware
failure.

Turn OFF meter and
turn it ON. Return to
dealer if necessary.

Err. 2 (in primary
display)

Memory checksum error

Hardware failure.

Turn OFF meter and
turn it ON. Return to
dealer if necessary.

Err. 3 (in primary
display)

ADC error

Hardware error.

Turn OFF meter and
turn it ON. Return to
dealer if necessary.

Err. 4 (in primary
display)

Keypad error

One or more keys on the
keypad are stuck.

Turn OFF meter and
turn it ON. Return to
dealer if necessary.

Err. Annunciator

Wrong keypad input

Wrong input in selected
mode.

Release key. Select
valid operations
depending on mode.

Electrode icon & buffer
icon blinks & Err
annunciator lights up

Calibration

Incorrect buffer used or
contaminated buffer
solution

Check if the right buffer
was selected or use
fresh buffer solution.

Battery icon lights up

Low Battery

Battery power is low

Replace batteries with
fresh ones as soon as
possible.