beautypg.com

Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 17

background image

13

NOTE: In pH 100 meter, you can program the meter to turn off the ‘READY’

indicator through the SETup mode.

10. Press the CON key to confirm the calibration. The ‘CON‘ indicator flashes for one second

and disappears.

11. Upon confirmation, the instrument is calibrated to the buffer indicated in the secondary

display.

12. The secondary display automatically scrolls to

the next pH buffer calibration option i.e. pH

10.01.

13. For 1-Point Calibration, this can be ignored and

you can exit to the [MEAS] mode by selecting

the CAL/MEAS key.

NOTE: The ‘Or’ indicator flashes if the selected

buffer value is not within +/- 0.50 pH

from the measured pH value.

14. To calibrate 2 or more points, press the Ml/

!

and MR/

"

key to scroll through the various

buffer pH options.

15. Select the second pH buffer for calibration

accordingly. Repeat steps 4 to 14 above.

16. Once calibration is confirmed at the second point,

you can proceed to calibrate at the third point

without the need to return to Measurement mode.

Just select the desired buffer by using Ml/

!

or

MR/

"

key.

NOTE: For pH 100 meter you can check on which buffer calibration points were being

calibrated using Program 2.2 under SETUP mode (see section 9.2.1 on Electrode Data).

NOTE: To exit from pH calibration mode without confirming calibration, DO NOT press

CON in step 11. Press CAL/MEAS instead.

If the selected buffer value is not within ±1.0 pH from

the measured pH value: the electrode and buffer icon

blink and the ERR annunciator appears in the lower

left corner of the display.

Figure 17: Err message and electrode

icon will appear if incorrect buffer are

used

Figure 16: Calibration for second point