Temperature calibration with atc probe – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 20

background image

16

4.4.2

Temperature Calibration With ATC Probe

1.

Make sure the ATC probe is attached to the meter.

Refer to Figure 7 in Section 3.2.2.

2.

Switch the meter on. Note the ATC indicator.

3.

Press the MODE key to select temperature mode,

[Temp].

4. Dip the ATC probe into a solution of known

temperature (i.e. a temperature bath). Allow time for

the temperature probe to stabilize.

5. Press the CAL/MEAS key to enter temperature

calibration mode. The primary display indicates the

pH value, and the secondary display shows the

temperature.

6. Press

!

or

"

key to select temperature.

7. Once you have selected the correct temperature,

press CON key to confirm.

8. The CON indicator flashes for one second and

disappears.

9. Press CAL/MEAS key to return to the pH

measurement mode.

NOTE: To exit this program without confirming the

temperature calibration value, DO NOT press CON.

Press CAL/MEAS instead.

Figure 19: Temperature calibration

with temperature probe