beautypg.com

Program 2 – electrode data, P2.0 : electrode offset, P2.1 : electrode slope – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 34: P2.2: calibration ph buffer data

background image

30

9.2

Program 2 – Electrode Data

This program allows you to check on electrode

parameters for diagnostic purposes.

9.2.1

P2.0 : Electrode Offset

The primary display shows the pH electrode offset value

in mV. The offset is calculated based on the buffer 7.00

calibration. If no calibrations have been performed, the

primary display shows -0.00 mV.

9.2.2

P2.1 : Electrode Slope

The primary display indicates the electrode slope in

percentage. The slope displayed is the average slope

based on the calibrations performed. If the meter has not

been calibrated, or under default conditions the primary

display reads 100.00.

9.2.3

P2.2: Calibration pH buffer data

Records all the pH buffer calibrations made on the meter.

This option allows you to view all 5 calibration points.

Use MI/

!

or MR/

"

key to scroll through the five

calibration points. See Figure 29.

If there has been no calibration performed at any

particular buffer option, the display will indicate “- - - -“.

See Figure 30.

Figure 29: View calibration points

Figure 30: No calibration performed

Figure 29: Electrode diagnostics,

offset & slope