beautypg.com

Program 1 – software initialization, P1.0 : memory clear, P1.1 : reset – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 33: Program 4 – communication setup

background image

29

9.1

Program 1 – Software Initialization

This program deals with the initialization of the memory

and calibration data. “OFF” is the default setting for both

Memory Clear and Reset. Accidental selection of wrong

option will wipe out the memory.

9.1.1

P1.0 : Memory Clear

Activation of this option by selecting ‘ON‘ clears all stored

measurement values in the memory (5 sets of values each

for pH, mV and Rel. mV with temperature). Under default

condition this option is not activated (set to OFF).

9.1.2

P1.1 : Reset

Activation of this option by selecting ‘ON‘, resets the

instrument. The instrument immediately switches off. You

need to power ‘ON‘ the instrument before proceeding with

any other functions.

RESET clears all data in memory (5 sets of values each for

pH; mV and Rel. mV with temperature), calibration data and

setup data. All settings return to factory defaults. The

RESET is OFF by default.

This feature is useful when the meter is used by someone

who needs to recalibrate to a different electrode and select

his own SETUP options.

Press CON key to continue to next program. NOTE: Data

cannot be selectively deleted.

Figure 28: Reset Meter

Figure 27: P1.0, Memory Clear