Ph calibration (without atc), Erasing calibrated values (for ph 100 only) – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 18

background image

14

4.3

pH Calibration (without ATC)

When the CyberScan meter is not used with a temperature probe, it is necessary to calibrate

the temperature first prior to pH calibration.

1.

Calibrate for temperature as described in the procedure in Section 4.4.1 - Temperature

Calibration Without ATC Probe.

2.

Next, proceed with the pH calibration as described in the above section.

3.

The meter is now calibrated at a fixed temperature.

NOTE: To exit this program without confirming the temperature calibration value, DO NOT

press CON. Press CAL/MEAS instead.