Temperature calibration, Temperature calibration without atc probe – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 19

background image

15

4.4 Temperature

Calibration

The temperature sensor is factory calibrated. Calibrate the

temperature probe only if you suspect the temperature

errors that may have occurred over a long period of time or

if you have a replacement temperature probe. This

procedure offers offset adjustment of probe to ensure more

accurate temperature measurement.

In the event where there is no temperature sensor, this

procedure can also be performed for Manual Temperature

Compensation (MTC).

4.4.1

Temperature Calibration Without ATC Probe

1.

Switch the meter on. Press the MODE key to select

temperature mode, [Temp].

2. Press the CAL/MEAS key to enter temperature

calibration mode. The primary display indicates the pH

value, and the secondary display shows the

temperature.

3. Press

!

or

"

key to select temperature value that

matches the temperature reading shown in your

reference thermometer.

4. Once you have selected the correct temperature,

press CON key to confirm.

5. The CON indicator flashes for one second and

disappears.

6. Press CAL/MEAS key to return to the pH

measurement mode.

NOTE: To exit this program without confirming the

temperature calibration value, DO NOT press CON. Press

CAL/MEAS instead.

Figure 18: Temperature

Calibration