7.2 Memory

Recall

This function recalls the previous readings stored in the memory. You can only access MR

from the measurement mode.

Make sure you are in the correct measurement mode to retrieve previously stored data. For

example, if you want to retrieve data stored during pH measurement mode, you will need to be

in pH measurement mode to retrieve the data.

To recall readings:

1.

Switch the meter ON, and press MODE key for

the appropriate measurement parameter.

2.

Press the MR key once to retrieve the last reading

stored.

3.

The memory annunciator, [MEM], will appear on

the display (See Figure 26).

4.

Note the reading.

5.

If necessary, press MR key again to retrieve the

next memory.

6.

To retrieve data from another measurement mode

(for example, mV), press MODE key until you

arrive at pH measurement mode.

7.

Press the MR key once to retrieve the last reading

stored in pH. Note the reading.

8.

If necessary, press MR key again to retrieve the next memory.

9.

Once all readings are recorded, press CAL/MEAS key to exit memory recall mode and

return to measurement mode.

NOTE: Memory data is retained even if the power of the unit is switched off. The memory will

be completely erased according to the following situation:

a.

When the batteries are removed (all functions, including memory, are reset);

b.

Memory can be cleared through SETUP function under program P1.0.

c.

When meter is reset through SETUP function under program P1.1.

Figure 26: Memory Recall function