Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 14

background image

10

Figure 13: Pin number position of the 9-pin RS 232C port

In case CyberScan pH 100 meter’s output has to be sent to a 25 pin RS232C connector, the

following cable configuration may be used:

CyberScan pH 100

25 pin connector

2 (TxD)

(RxD) 3

4 (DSR)

(DTR) 20

5 (GND)

(GND) 7

7 (CTS)

(RTS) 4

1.

CyberScan pH 100 uses hardware handshake i.e. CyberScan pH 100 expects both DSR

and CTS lines to be active before it sends data.

2.

If the [PRINT] key is pressed while the printer is not ready or if the printer is off, the

CyberScan pH 100 meter displays error message by showing both the printer and the

ERR annunciators blinking alternately, and awaits the printer to be ready (Figure 14).

3.

While the meter is displaying printer error, you

may press CAL/MEAS key to return to the

measurement mode.

Figure 14: Error in printing