20

5.2

Manual Temperature Compensation

IMPORTANT: For manual compensation, you must

disconnect the temperature probe. The [ATC] annunciator

will disappear from the LCD.

1.

Switch the meter on. Press the MODE key to select

[TEMP] mode.

2. Press the CAL/MEAS key to enter Temperature

calibration mode. The CAL indicator will appear above

the primary display.

3.

The primary display shows the current pH reading and

the secondary display shows the default temperature

value 25 ºC.

4. Check the temperature of your sample using an

accurate thermometer.

5. Press

the

"

or

!

keys to set the temperature to the

measured value obtained from step 4.

6.

Press CON to confirm the selected temperature. The

[CON] indicator will flash momentarily.

7. Press the CAL/MEAS key to exit and return to pH

measurement mode.

The meter will now compensate pH readings for the

manually set temperature.

NOTE: To exit this program without confirming the manual

temperature compensation value, DO NOT press CON in

step 6. Press CAL/MEAS instead.

Figure 22: pH measurement using

Manual Temperature

Compensation