MEMORY FUNCTION (FOR pH 100 ONLY)
The memory function is available only in the CyberScan pH 100 meter. This meter stores data
in sets:
•
pH and temperature
•
mV and temperature
•
Relative mV and temperature
The pH 100 meter can store up to 16 sets of data in any combination of values. For example,
you can store 7 pH and 9 mV values.
This meter uses the Last-In-First-Out (LIFO) method of memory management. If the memory
is already full, the first value that was stored in the memory will be erased to create space for
the new value to be input.
7.1 Memory
Input
To store a reading:
1. During any measurement function [MEAS], press
MI key to input any data into the memory.
2. The [MEM] annunciator will flash momentarily
(Figure 25).
3.
The meter then returns to measurement mode.
4.
You can switch to the next measurement mode (by
pressing the MODE key), and press MI key to input
data into memory.
NOTE: If the memory is full, the first value stored will be
erased to create space for the new value.
Figure 25: Memory In function