Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 54

background image

50

g) Test with standard calibration buffer solutions.

CAUTION: Proper eyewear and gloves must be worn when preparing these chemicals and

performing this operation.

Finally, test with standard calibration buffer solutions to see if the electrode yields acceptable

results. You may repeat steps ‘c’ to ‘f’ again for better response (maximum 3 times). If the

response does not improve, then the electrode has completed its useful life. Replace with a

new electrode.