beautypg.com

Electrode care, Electrode maintenance, Storage – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 52: After use, Electrode activation, Rejuvenation procedure, 11 electrode care

background image

48

11 ELECTRODE

CARE

11.1 Electrode

Maintenance

pH electrodes are susceptible to dirt and contamination and need to be cleaned regularly

depending on the extent and condition of use.

11.1.1 Storage

The best results, always keep the pH bulb wet, preferably in pH 4 buffer with 1/100 part of

saturated KCl. Other pH buffers or tap water are also acceptable storage media, but avoid

storage in de-ionized water. The protective rubber cap filled with the buffer solution provides

ideal storage for long periods.

11.1.2 After

Use

After measurement is complete, follow the sequence elaborated below for storage.

1.

Wash the electrode and reference junction in deionized water.

2.

Close the refilling hole by returning its rubber sleeve or stopper cap (Necessary for only

refillable electrodes).

3.

Store the electrode as mentioned above (Storage section 11.1.1).

11.1.3

Electrolyte Replacement (for refillable electrodes only)

The reference electrolyte needs to be refilled when the electrode has been used for an

extended period, or when the internal electrolyte has dried up. To accomplish this, follow the

procedure detailed below.

a) Remove the protective rubber cap or sleeve

Remove the protective rubber sleeve to expose the filling port of the electrode. Remove the

old reference electrolyte with a syringe and a flexible tube.

b) Fill the new reference electrolyte

Add in fresh electrolyte until it reaches the level of refilling port. The reference electrolyte used

should be 4M KCl. Replace the rubber sleeve.

c) Re-use the electrode

Rinse the liquid junction with deionized water and tap dry.

NOTE: If these steps fail to restore normal electrode response, you may attempt to rejuvenate

it. (See section 11.4 on Rejuvenation Procedure).