beautypg.com

Insert the electrode into the holder, Connecting the ac/dc adapter, 4 calibration – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 12

background image

8

3.2.5

Insert the electrode into the holder

1.

Do not use excessive force when inserting electrodes into the holders.

2.

Insert the pH electrode into the opening of the first holder until the top housing of the

electrode touches the top of the holder.

3.

If you are using a separate temperature probe, insert the probe into the opening of the

second holder until the ridge on the housing touches the top of the holder.

NOTE: The holder is designed for probes 12 mm in

diameter. Electrodes larger than 12 mm may not fit in

the holder. Forcing the electrode into the opening

may damage the holder or your electrode.

The electrode holders can be attached in different

positions for greater flexibility in measurement and

storage purposes. Simply slide out the electrode

holders and reoriented into appropriate

orientation before putting into position.

3.3

Connecting the AC/DC Adapter

Besides using four AAA-sized batteries as

power source, the CyberScan pH 10/100

meter can also operate from the power

mains using an AC/DC power adapter either

at 120/220 VAC (sold separately). This is

extremely useful if you have an A.C. power

source available (e.g. laboratory).

Before plugging in, switch off the meter and

the power source of adapter. This is a safety

precaution that should be adhered to

safeguard your CyberScan meter.

1.

Switch off the meter and power sources.

2.

Select the correct voltage of AC/DC Adapter. See Figure 11.

NOTE: Output Voltage: 12 V D.C.; Current: 500 mA. Ensure that the input mains

voltage (120/240 VAC) matches your adapter requirements.

3.

Insert the D.C. jack into the socket and switch on the power to the adapter, followed by

the CyberScan meter.

Figure 11: Putting the AC Power Adapter

Figure 10: Different positions using

electrode holder