beautypg.com

Ph calibration with atc – Thermo Fisher Scientific CyberScan pH 10/pH 100 User Manual

Page 16

background image

12

4.2 pH

Calibration

with

ATC

To activate the Automatic Temperature

Compensation (ATC), simply plug in the

temperature probe into the phono jack. The ATC

indicator will be displayed on the LCD (Figure

15). If the ATC indicator is not displayed, it

indicates that the temperature probe is either not

properly connected to the instrument or is faulty.

1.

Switching ‘ON’ the meter by pressing the

ON/OFF key.

2. All the LCD segments will be displayed

momentarily for a few seconds to check

that all segments are in working condition.

Then, the LCD will switch to the [pH]

measurement mode.

3.

Select the [pH] mode using the MODE key.

(See Figure 15 on right).

4. Rinse the electrode well with deionized

water or rinse solution. (Do not wipe the

electrode with tissue paper as this may

cause a build-up of electrostatic charge on

the glass surface!).

5. Dip the probe into the calibration buffer.

The end of the probe must be completely

immersed into the sample. Stir the probe

gently to create a homogeneous sample.

6.

Press the CAL key to calibrate the meter.

7.

The display will show the [CAL] mode and

the buffer icon as shown (see [1] & [2] in

Figure 15).

8. The primary display will show the

measured reading while the secondary

display will indicate that pH 7.00 is ready for

calibration.

9.

Wait for the measured pH value to stabilize.

The READY annunciator will light up. See [3]

in Figure 15.

Figure 15: pH Calibration at pH 7.00