beautypg.com

SHIMPO DT-601CG User Manual

Page 45

background image

15. ADJUSTING ANALOG OUTPUT

(AV/AI OPTION)

Specification(AV3-5/AI) of product is being adjusted accurately by us.
Do not touch excluding the emergency.

Methode of adjustment

(1) Call test mode by turning on while pushing Mode key.

(2) Call the analog output test by pushing Mode key at few times.

( Refer to "Setting menu" on page15 )

(3) Adjust "Span-volume/Zero volume" to become the following values.

(Adjust Span-volume first.)

Voltage output : AV3 type

Voltage output : AV4 type

Voltage output : AV5 type

Voltage output : AI type

Display

Voltage

0

1V

10

5V

Turn "Zero volume"
Turn "Span volume"

Display

Voltage

0

0V

Turn "Zero volume"

10

5V

Turn "Span volume"

Display

Voltage

0

0V

Turn "Zero volume"

10

10V

Turn "Span volume"

Display

Voltage

0

4mA

Turn "Zero volume"

10

20mA

Turn "Span volume"

Zero volume

Span volume

P46