13. SETTING OF DISPLAY OFFSET
Setting of display in reset. Such as when setting of display offset "001000", the display in reset is "1000"
and measurement begin from "1000". If you begin to measurement from "0", setting of display offset "000000".
The following table refer to key operation to setting of display offset.
The setting range is "-99999 - 999999".
(Setting of the decimal point is connected to [Mode 00] in P19.
Initial value is "000000"
Caution
1. Do not turn off the power while registering the set value
(from pressing ENT until return to measurment display)
2. Mode Protect does not function.
3. Reset it before starting measurement when Display Offset Value is changed.
Display
Operation
Operation Key
Press Mode key and Up key more than 2seconds.
D1LED's LED light, then Display Offset of D1 is
called.
PRESET Value is switched pushing Mode Key.
Also, The indicator-A displays "-".
Change to right the position of flashing digit.
One hit makes 1digit move right hand side.
Change the value of flashing digit.
One hit makes the value bigger/smaller by 1.
Register the set value. After finishing the
setting,register setting with this key. After
registration, measurment display will appear.
Return to measurement display without registering
the set value.
Push reset key after register to the display offset
value, and display the setting of display offset.
Measurement begin from this value.