beautypg.com

KROHNE Summit 8800 Vol 3 User Manual

Page 33

background image

33

www.krohne.com

08/2013 - MA SUMMIT 8800 Vol3 R02 en

SUMMIT 8800

DATA LOGGING

05

Figure 21 Data log settings

Read access level

An access level for reading these data (for SOAP protocol only)

Write access level

An access level for writing these data (for SOAP protocol only)

Setup name

A name for this data log

Log every

If the log is periodically, set here the log interval between 5 minutes and 1 year

Num. records

The number of records after which the log start from the beginning again

Log change of

If the log is event driven, select which variable, from the ID tree, triggers the

event
Be careful to select an variable which changes state, such as end of prove

Default display page

Press when page must be displayed after the screen is not used for a period of

time

Add to display menu

If checked, this mimic will be placed in the menu item “data logs”
If not checked, the mimic can be used as a display page.

The statistics of memory used is indicated on this page, in terms of time and amount of memory

used.

Figure 22 Data log statistics

The period of time the total record will cover

• How much memory is used for this log record

• How much memory is used for all log records

• How much memory is still free

• The maximum amount of memory available for logging.

Please note:

• All data log items are stored in either internal data flash memory or removable SD card

memory depending upon the configuration. Each record contains a CRC 32 which is generated

when the log record is created and checked for validity each time data is read from the Unit.