9 .3 Translation
Liquid
Gas
Steam
English is the default language for the Summit, but it is possible to define a new language.
9 .4 Audit log
Liquid
Gas
Steam
For diagnostic purposes, the user can extend the audit trail with alarms and with up to 10 ad-
ditional data items per event. These data will be stored on an external SD card only.
9 .5 Settings
Liquid
Gas
Steam
General calculation settings are used to set the cycle time (how often all measurement and cal-
culations must be done per second), the maximum counter value and what to do with negative
station flow:.:
Figure 209 General settings
Calculation time
The cycle time or the time after which the Summit will start again
with all calculations. This number should be larger than the current
calculation duration. At the maximum, the Summit will do 4 full
cycles per second.
Counter rollover limit
The maximum counter value after which the counter restarts from
zero.
Allow negative station counters
If not, then a negative station flow will be counted as zero.