SUMMIT 8800

COMMUNICATION

08

For a Ethernet port set the following parameters:

Enable

Click the box to enable modbus over TCP/IP

Port

Ethernet port number

ID

The modbus identifier for this link (1..255)

Name

Any descriptive name to identify communication port

8 .4 .1 .2 Modbus timeout setting

In the settings, it is also possible to enable the modbus timeout to create an alarm when the

counter exceeds the number of seconds set under the general Modbus Timeout:

Figure 179 Modbus slave enable timeout

Check the box is the modbus timeout may be used. This setting refers to a general setting for

modbus timeout:

Figure 180 Modbus slave timeout settings

Modbus Communication is subject to 3 re-tries and a 1-second timeout on each re-try. If at the

end of this no reply or an invalid reply is received then an alarm appropriate to the failure is is-

sued. The Modbus time-out function is designed to identify when any essential Modbus master

or slave communication is lost.

To disable the timer, set the timeout value to -1 otherwise enter the maximum time that the

SUMMIT 8800 is allowed to be without Modbus communications. An accountable alarm will be

raised on every stream that is using the time-out alarm if communication loss occurs longer

than the defined time allowed. All Modbus received gas or liquid data will be flagged as an error.