beautypg.com

GW Instek GSP-827 User Manual User Manual

Page 71

background image

2.7GHz Spectrum Analyzer Operation Manual

70

Insert a point

0

99.5

-45.0

5 100.1 -45.0

1

99.7

-45.0

6 100.4 -45.0

2

99.9

-45.0

7 100.5 -45.0

3

99.9

-30.0

4

100.1

-30.0

When a new point is inserted at its current field, the temporary values of frequency and level are
automatically generated. Both new values are the average of ahead and followed points.

Enter the desired frequency. Take 100.3MHz as example here.

0

99.5

-45.0

5 100.1 -45.0

1

100.3

-45.0

6 100.4 -45.0

2

99.9

-45.0

7 100.5 -45.0

3

99.9

-30.0

4

100.1

-30.0

When the entry is completed, the table will sort itself immediately by the order of frequency. The
limit line is modified accordingly. Enter the Level value if it is necessary.

0

99.5

-45.0

5 100.3 -40.0

1

99.9

-45.0

6 100.4 -45.0

2

99.9

-30.0

7 100.5 -45.0

3

100.1

-30.0

4

100.1

-45.0

Limit Line

Edit

Insert

0

1

2

3

7

6

5

4

8

9

.

-

GHz

dBm

MHz

dBmV

kHz

dBuV

BK SP

Start:99.5MHz

Stop:100.5MHz

VBW:100 kHz Center:100.0MHz

Span:1.0MHz

Ref

Lvl: -30.0 dBm

-30.0

-50.0

-70.0

-90.0

-110.0

SAT JAN 28, 2003

14:30:25

INT:REF

AMPL CAL

Limit

Lower

Insert

Undo

Delete

Reutrn

Frequency

Frequency

Level

Level

(dBm

(dBm

(MHz)

(MHz)

0: 99.5

-45.0

1: 99.9

-45.0

2: 99.9

-30.0

3: 100.1

-30.0

4: 100.1

-45.0

5: 100.3

-40.0

Level:-45

6: 100.4

-45.0

7: 100.5

-45.0

1