2.7GHz Spectrum Analyzer Operation Manual

42

Input Z 50

Ω/ 75Ω

Keep pushing Input Z key will switch the input impedance between 50 and 75Ω. The prompted
message will appear that the “zz adaptor” should be connected.

Input Z Cal

UP/DOWN

and Scroll: To change offset by 0.1dB for every move.

Editing

: Editing keys can adjust the offset from -10.0 to 10.0dB in 0.1 dB resolution. The ideal

number is at 5.9dB

The “Input Z Cal” performs only when the input Z is at 75Ω. In other words, when the input Z is
switched to 50Ω, the calibration value will return to 0.

Amplitude

Input Z

50Ω

1

Start:99.5MHz

RBW:30 kHz

Stop:100.5MHz

VBW:100 kHz

Center:100.0MHz
Span:1.0MHz

Sweep:100.0ms

75 Ohm adaptor

Ref

Lvl: 77.0 dBuV

77.0

67.0

57.0

47.0

37.0

27.0

17.0

-3.0

7.0

SAT JAN 28, 2003

14:30:25

INT:REF

AMPL CAL

Ref

Level

InputZ

75Ω

Scale

5dB

Unit

dBuV

ExtGain

Loss

InputZ

Cal

1

Amplitude

Input Z

Cal

Start:99.5MHz

RBW:30 kHz

Stop:100.5MHz

VBW:100 kHz

Center:100.0MHz
Span:1.0MHz

Sweep:100.0ms

Input Z Loss:5.0 dB

Ref

Lvl: 82.0 dBuV

82.0

72.0

62.0

52.0

42.0

32.0

22.0

2.0

-95.0

SAT JAN 28, 2003

14:30:25

INT:REF

AMPL CAL