2.7GHz Spectrum Analyzer Operation Manual

20

F6

Setup>

Set the frequency of trigger source. The default is Center

frequency.

N, … MHz

Key in the frequency.

F6

Exit the trigger setup submenu.

F4

Select single trigger, there will be only one sweep..

Otherwise, it will have continuous sweeps.

F5

Delay>

Define delay between the trigger condition and sweep.

F2

Level >

Exit the Single trigger mode.

F1

Exit the trigger mode.

6.2.15 Trigger by Externally stimulus signal

Function key

Description

Trigger

Select the trigger mode functions.

F3

Select the external trigger input (on the rear panel) as the

trigger source.

Note: The Single trigger, trigger delay, trigger setup and exiting single trigger

operations are same as previous “Video Level” mode.

6.2.16 Observe two sweeps (ex, 2

nd

harmonic) by Dual Windows

Function key

Description

Display

Select the Display relative functions.

F4

Split the window in dual windows.

F5

Select the upper window as active one. All the measurement

configurations are due to upper one. In harmonic

measurement example, set the fundamental frequency as the

center frequency here.

F6

Select the lower window as the active one. All the

measurement configurations are due to lower one. In

harmonic measurement example, set the harmonic

frequency as the center frequency here.

6.2.17 Configure Display setting

Function key

Description

Display

Activate the power measurement relative functions.

F1

Cntrst >

Rotate Spinner to adjust LCD contrast.

F2

Switch the backlight ON/OFF.

F3

screen>

Invert the screen.