2.7GHz Spectrum Analyzer Operation Manual

9

4.0 First Time Use

The following text assumes that the SAFETY section of this manual has been read carefully

and understood.

Each time before the instrument is operated, make sure that it is connected to protective earth.

Before establish the connection and measurement, check if the device under test is connected

to protective earth as well.

Since the RF measurement is very sensitive, clean connectors on the front panel is very

important for the accurate measurement. Regularly cleaning up the connector is strongly

recommended.

The main power switch is located on the rear side. Switch it on to activate the Standby mode

and the power indicator on the front panel is on with red light. Push and hold the “STBY”

button for 2~3 seconds to turn on the unit. The power indicator will turn into green. Pushing

and holding the same button turns the unit to standby mode. When the unit is turned into

standby or off, the last setting is stored and recalled when it is turned on again.

4.1 Internal calibration signal

A 100MHz -30dBm calibration signal is built-in for internal calibration. Turn it on and off by
the following function keys.

Sysetm : Push System key.

F3 : Push F3 key to switch the internal calibration signal ON/OFF

.

It is not a well-filtered signal. When the unit is turned on, if 100MHz and harmonics appear on
the display, check this signal is ON or OFF first.