beautypg.com

GW Instek GSP-827 User Manual User Manual

Page 24

background image

2.7GHz Spectrum Analyzer Operation Manual

23

F3

Config>

The system configuration will be shown up, including serial

number, S/W and F/W version number, Options, Los status,

reference signal source and status.

F6

Exit the submenu.

6.2.24 Synchronized by External Reference Signal

Function key

Description

Option

Select the Option relative functions.

F3

Freq>

Select the external reference signal source.

Up/Down

Arrows

or

Spinner

Select the reference signal frequency.

-

Connect the reference signal source to the BNC terminal

“Ref Input” on the rear panel.

6.2.25 Synchronize other equipments

Connect the “10MHz Ref Output” terminal on the rear panel to the “External Ref. Input” of

following equipment directly.

6.2.26 Tracking Generator Operation (option)

Function key

Description

Option

Select the Option relative functions.

F1

Select the Tracking Generator.

F2

>

Set the TG output Level.

F1

Switch the TG ON.

F3

Perform the Normalization function before connecting the

DUT if necessary.

F5

Value>

This function allows normalization on the specific level for

extra gain or loss.