2.7GHz Spectrum Analyzer Operation Manual
⎯
⎯
40
Reference Level
UP/DOWN and Scroll
: To change the Reference Level one scale a time. If the scale is at 10dB,
the UP key will increase the RefLvl from –30dBm to –20dBm in the above case.
Numerical
: Using the numerical and unit keys can specify the reference level from –50 dBm to
+20dBm in one dBm resolution.
Scale
UP/DOWN
and Scroll keys switch the scale in 10-5-2-1 sequence.
Editing
: The editing entry will select the closest scale to 10-5-2-1 one.
1
Start:99.5MHz
RBW:30 kHz
Stop:100.5MHz
VBW:100 kHz
Center:100.0MHz
Span:1.0MHz
Sweep:100.0ms
Ref Lvl:-20dBm
Rel
Level
Ref
Lvl: -20.0 dBm
-30.0
-40.0
-50.0
-60.0
-70.0
-80.0
-90.0
-20.0
-100.0
SAT JAN 28, 2003
14:30:25
INT:REF
AMPL CAL
Scale
10 dB
Unit
dBm
ExtGain
Loss
Amplitude
Ref
Level
0
1
2
3
7
6
5
4
8
9
.
-
GHz
dBm
MHz
dBmV
kHz
dBuV
BK SP
Input Z
50Ω
Input Z
Cal
1
Amplitude
Scale
5dB
Start:99.5MHz
RBW:30 kHz
Stop:100.5MHz
VBW:100 kHz
Center:100.0MHz
Span:1.0MHz
Sweep:100.0ms
Scale:5 dB/Div
Ref
Lvl: -30.0 dBm
-30.0
-35.0
-40.0
-45.0
-50.0
-55.0
-60.0
-70.0
-65.0
SAT JAN 28, 2003
14:30:25
INT:REF
AMPL CAL
Ref
Level
InputZ
50Ω
Scale
5dB
Unit
dBm
ExtGain
Loss
0
1
2
3
7
6
5
4
8
9
.
-
GHz
dBm
MHz
dBmV
kHz
dBuV
BK SP