Emo‐3000 – AW Gear Meters EMO-3000 User Manual
Page 39

EMO‐3000
Operation and Programming Manual
38
Transmitter
KFT cc
KFR Gal
KFR Oz
ZHM‐01
236.6
.0624
8.00
ZHM‐02/1
1200.
.319
40.84
ZHM‐02
2271.
.599
76.66
ZHM‐03
5746.
1.518
194.0
HPM‐15
1200.
.319
40.84
HPM‐20
2271.
.599
76.66
HPM‐30
5746.
1.518
194.0
As these are generic K‐factors, consult the calibration sheet supplied with the flow transmitter for the
most accurate information.
USED IN:
All modes.
RELATED VARIABLES:
DP for Rate A KFT (only in OPTO communication).
HEX ADDRESS:
00E4 HEX ‐‐‐‐> same address used in EMO‐2000.
BYTES:
2 Bytes DEFAULT: 1200
ASC form COMMAND:
reading = 107; writing = 507; E is with decimal point. N is just the
numeric value.
Engineering units for KFR
The KFR units is stored as a three‐letter string. This means any engineering unit can be constructed if a
computer hookup is used.
USED IN:
All modes.
RELATED VARIABLES:
KFT Engineering units.
HEX ADDRESS:
0052 HEX ‐‐‐‐> same address used in EMO‐2000.
BYTES:
3 Bytes DEFAULT: cc/m
ASC form COMMAND:
reading = 109; writing = 509; E is engineering.
Engineering units for KFT
The KFT units is stored as a three‐letter string. This means any engineering unit can be constructed if a
computer hookup is used.
USED IN:
All modes.