Page 28

background image

EMO‐3000

 

Operation and Programming Manual 

 

 

 

27 

      
 

 

Use F2 = PGUP in the LIMIT VARIABLES screen to access the following variables:  

 FLOW  LIMIT  2 =  

XXXX.X 

 FLOW  MARGIN 2 = 

XXXX.X 

 FLOW  LIMIT  3 = 

XXXX.X 

 FLOW  MARGIN 3  = 

XXXX.X 

 FLOW  LIMIT  4 = 

XXXX.X 

 FLOW  MARGIN 4 = 

XXXX.X 

When the desired variable is moves to the arrow, select F4 = PROGRAM. The DM‐3000 then asks for the 
pass code.   

ENTER PASS CODE = XXXX 

The code is 4146.  

Key in the correct variable value and Enter with F6. 

The Limit Variables work in conjunction with the Limit Rules in the next section. The values entered 
under the Limit Variables determine the trip points of any limit programmed to work as a function of 
them. For example, if Total Limit 1 in the Limit Variables section above are programmed to 1000, and 
the Limit Rule discussed in the next section of this manual is programmed to function with Total Limit 1, 
the limit output turns ON only when the totalized fluid value exceeds 1000 counts in the engineering 
units programmed (gallons, ounces, CC’s, etc.). Therefore, both the Limit Value and the Limit Rule are 
needed for the desired action to occur. 

1 = LIMIT RULES 

     * * * LIMIT  RULES        * * *  CH = XXX 

‐‐‐‐>   LIMIT 1 RULE 

=   000XX. 

 

LIMIT 2 RULE 

=   000XX. 

 

LIMIT 3 RULE 

=   000XX. 

 

LIMIT 4 RULE 

=   000XX.           

 

RATIO 1 +/‐% 

=   XXXXX. 

 

RATIO 2 +/‐% 

=   XXXXX. 

F1 = ESC 

F2 = PGUP   

F3 = PGDN   

F4 = PRGM 

 

        

The last two variables in the screen above are not related to the LIMIT RULES. They appear as part of the 
variable listing only by coincidence.  

Notice 

Notice