beautypg.com

Emo‐3000 – AW Gear Meters EMO-3000 User Manual

Page 14

background image

EMO‐3000

 

Operation and Programming Manual 

 

 

 

13 

      
 

ma‐Scaler  

The milliamp scaler scales the 0 ‐ 20 mAmp input signal on Pin 5 and 6 to the flow rate in 

engineering units/mA.  

For example: 20mA = 1000 cc/min. Then,  

1000 

 

 = mA‐Scaler 

where B = 20mA ‐ mA offset value.  

With a mA Offset Value of 4mA, then mA‐Scaler = 62.5 cc/min or 1mA = 62.5 cc/min. 

This signal serves as a flow rate setpoint in the PID mode after scaling. 

mA’s offset  

Used  in  a  situation  where  there  is  not  an  exact  4  mA’s  at  the  low  point  of  the  analog 

input. The offset is the base line for calculating the set‐value based on the analog input. 

Important Hardware Connections for PID Operation

None of the following external signals has to be wired. However, they are useful features in PID mode.  
See Appendix P for details. 

Transparent  

The transparent input redirects the analog input, (PINS 5/6) directly to the mAmps 

output (PIN 7). It is necessary to use the transparent input if the channel is put in open loop simulation.  
This signal is located at PIN 11 on the backplane. 

Hold Analog  

This input signal holds the analog output at the last set output. The hold input is often 

used in flush situations where the controller is supposed to be disabled. This signal is at PIN 12. 

Hold total   

Hold total input disables the totalizer. Often used during flush or soft‐air push cycles. 

This signal is at PIN 13 on the backplane.    

Set‐value  

You must program this number in order to generate a flow rate in PID if the external 0‐

20 mAmp set value is not hooked up. This is the digital set‐value. It is important to know that the analog 
input set‐value is overridden. 

Set reached    An output that activates when the flow rate meets the set‐value (+/‐ tolerance). This 
signal is at PIN 14.  

Fluid Tables:    The EMO‐3000 supports 32 fluid tables. Each table is externally accessed via pin #8 or 
via the DM‐3000.  The default fluid number is 00.  Most variables affect all 32 fluid numbers, but a few 
require that you enter the fluid number when they change. In this event, an additional screen appears 
with prompts. You can assign each fluid table to store flow data for a different fluid or color as it is being 
regulated. These data tables are continually adapted as each color is encountered. Each fluid is 
identified by a voltage of 0.15625 volts, which represents 5 V divided by 32. For example, the first 
color/fluid is indicated by the step voltage 0‐0.15625 V. See the fluid table illustration on page 22.